Hospitals use radiation in X-ray and CT rooms, radiotherapy and nuclear medicine. An area radiation monitor is a fixed instrument that measures the dose rate at the point where it is mounted. It shows the people responsible when levels change, so they can act sooner. However, it is one part of a radiation safety programme, not a replacement for it.
Key takeaways
- Risks differ between X-ray, radiotherapy and nuclear medicine, so monitoring needs differ too.
- Levels outside iodine-131 rooms vary by hospital, position and time.
- An area radiation monitor measures one point. It does not measure what a person receives.
- Alarm thresholds come from the RSO or medical physicist, not from annual dose limits.
Where is there radiation in a hospital?
Diagnostic radiology. An X-ray unit produces radiation only while it is switched on. When the exposure ends, no radiation remains in the room. Shielding protects nearby areas, and it is checked by survey when the room is commissioned [1].
Radiotherapy. Treatment rooms need heavy shielding, a maze entrance, warning lights and a door interlock that stops the beam [1]. Brachytherapy uses real radioactive sources. For high dose-rate (HDR) sources, the Office of Atoms for Peace (OAP) requires a room monitor that shows when the source is out. Staff also confirm with a survey meter that the source is back in its shielded position [2].
Nuclear medicine. Here the sources are unsealed, so they cannot be switched off. Patients who receive a radiopharmaceutical emit radiation and excrete radioactive material. For this reason, the OAP says contamination must always be checked, especially in the lab and the radioactive material store [2].
What studies of iodine-131 rooms show
Rooms for patients treated with high-activity iodine-131 (I-131) have the most published measurements. These studies used dosimeters or survey meters, not area monitors. Still, they show what a monitoring plan must account for.
- Thailand: In a non-lead-lined children’s room with mobile lead shields, the adjacent wall and corridor read 1.82–4.48 µSv/h. The study compared this with a 0.5 µSv/h public level [3].
- United States: With rolling lead shields, the mean maximum outside the room was 0.013 mSv/h, and staff and parent doses stayed below limits [4].
- Jordan: The corridor was near background. However, estimated annual doses at the patient’s bed and in the bathroom were above the worker limit [5].
The Thai study’s largest dose was inside the room. A parent sitting by the bed received up to 7,433 µSv [3]. A monitor outside the room would not show this, so caregivers need personal dosimeters and clear guidance.
Levels also change over time. A 2025 Australian study used ceiling-mounted detectors that measured every minute. Across 250 patients, it tracked how quickly each patient’s radiation fell [6]. In short, results differ widely between hospitals, so figures from another site cannot replace your own measurements.
Area radiation monitor vs personal dosimeter
| Area radiation monitor | Personal dosimeter | |
|---|---|---|
| What it measures | Dose rate at the one fixed point where it is mounted | Dose received by the person wearing it |
| What it is for | Showing the people responsible when levels in an area change | Tracking each person’s dose, including caregivers by the bed |
| What it does not show | What a person receives, or surface contamination | Levels in an area when no one is wearing it there |
An area radiation monitor also does not detect surface contamination. A Japanese study checked an I-131 room for three weeks after each patient left. It found negligible risk to staff once patients had gone [7]. A Chinese study found contamination below limits everywhere except the toilet [8]. Therefore, contamination checks remain a separate task.
Setting alarm thresholds
ICRP Publication 103 limits workers to 20 mSv per year, averaged over five years, and the public to 1 mSv per year [9]. In Thailand, the Nuclear Energy for Peaceful Purposes Act and the Ministerial Regulation on Radiation Safety B.E. 2561 apply these standards, with the OAP as regulator [10] [11].
These are yearly limits for people, not alarm levels for rooms. So the RSO or medical physicist sets each threshold from the area’s risk assessment, normal background and how people use it.
Where to install an area radiation monitor
Positions should come from a risk assessment, not from where cables are easy to run. Common examples include:
- Inside an HDR brachytherapy room, to show when the source is out [2].
- The anteroom to I-131 patient rooms [12].
- The radioactive waste store [12].
- Above the I-131 patient bed, to read the patient’s dose rate remotely [6] [12].
- A corridor or adjacent wall, only where the risk assessment shows a need.
Before installing, also check that the instrument suits the radiation type, that alerts reach named staff at night, and that calibration follows the manufacturer and regulator.
How Sitearound connects area radiation monitors
Sitearound installs and connects each area radiation monitor to the Sitearound IoT platform, so readings from every device appear on one screen. We use [detector brand and model], type-tested to [standard] and calibrated by [calibration lab].
When a reading reaches a threshold set by the hospital’s RSO, the system alerts named staff by email or messaging app. It also opens a work order, so you can see who responded and when. Historical data can be exported for safety reports and inspections.
Note: The system handles data and coordination. Safety decisions stay with the hospital’s radiation specialists.
References
- International Atomic Energy Agency. Radiation Protection and Safety in Medical Uses of Ionizing Radiation (SSG-46). Vienna: IAEA; 2018.
- Office of Atoms for Peace (Thailand). Radiation Safety in Medical Practice [in Thai].
- Chuamsaamarkkee K, et al. Radiation monitoring of non-lead-lined treatment room in general pediatric ward and adjacent areas for high dose 131Iodine-mIBG. Polish Journal of Medical Physics and Engineering. 2018;24(3):133–136.
- Chu BP, et al. Feasibility of administering high-dose 131I-MIBG therapy to children with high-risk neuroblastoma without lead-lined rooms. Pediatric Blood & Cancer. 2016;63(5):801–807.
- Abu-Khaled YS, et al. Radiation exposure from radioactive iodine 131I in and surrounding the patients’ room. Radiation Protection Dosimetry. 2009;135(1):64–68.
- Continuous monitoring of radiation emissions from 131I thyroid cancer ablation subjects (250 subjects). Asia Oceania Journal of Nuclear Medicine and Biology. 2025.
- Karo C, et al. Radiation monitoring of an isolation room for 131I therapy after the patients were released. Health Physics. 2019;117(4):419–425.
- Investigation on the radiation of 131I in treatment site of a grade A tertiary hospital [in Chinese]. Chinese Journal of Radiological Health. 2022.
- ICRP. ICRP Publication 103: Dose limits. ICRPaedia.
- Nuclear Energy for Peaceful Purposes Act B.E. 2559 (2016), as amended by (No. 2) B.E. 2562 (2019). Royal Gazette. [Add Royal Gazette link]
- Office of Atoms for Peace (Thailand). Ministerial Regulation on Radiation Safety B.E. 2561 (2018) [in Thai].
- Chia KL, et al. A targeted radionuclide therapy facility at a university teaching hospital in Singapore. Cureus. 2025;17(9):e92308.